A study on Accuracy of a Reusable Probe and Regional Oximeter Based on Near-Infrared Spectroscopy During Controlled Hypoxia in Healthy Participants

Lijuan Tang, Yufeng Zou, Mengyun Li, Shangrong Li*, Zongze Zhang*

Abstract


Objective: Regional cerebral oxygen saturation (rSO2) has been widely used in various clinical settings. However, present oximeters on the market are disposable and expensive, and their accuracy is various. We conduct an accuracy study of a reusable probe and regional
oximeter developed by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
  Approach: The regional oximeter was connected and the supporting reusable near-infrared spectroscopy sensor was placed on 26 healthy
volunteers' forehead for rSO2 measurement. The participant's inspired oxygen was changed according to the 10 set platforms. Simultaneous
blood samples from the jugular bulb and radial artery were taken and analyzed for reference SavO2. Comparing the rSO2 (monitored by the
regional oximeter) to the reference SavO2, the absolute and trend accuracy of the regional oximeter were calculated. Multiple linear regression
analysis was performed on the effects of sex, skin color, and age.
  Main Results: The bias and standard deviation of absolute accuracy was 0.80% and 2.90%. Trend accuracy analysis yielded a bias of
0.92% with a standard deviation of 2.53%. The root-mean-square accuracy of absolute accuracy (Arms) was 3.00%; the root-mean-square accuracy of trend accuracy (Arms-t) was 2.69%; the Arms and Arms-t under SavO2 ≤ 60% are 2.98% and 2.82%; the Arms and Arms-t under SavO2 > 60%
are 3.01% and 2.55%. Gender, skin color and age did not significantly affect the accuracy (all P > 0.05).
  Significance: rSO2 readings by Mindray's device and the reusable probe had good absolute and trend accuracies with respect the calculated reference values.

Keywords


Regional cerebral oxygen saturation; Near-infrared spectroscopy; Reusable probe

Full Text:

PDF

Included Database


References


[1] Huppert EL, Parnia S. Cerebral oximetry: a developing tool for monitoring cerebral oxygenation during cardiopulmonary resuscitation.

Ann N Y Acad Sci. 2022;1509(1):12-22.

[2] Takegawa R, Hayashida K, Rolston DM et al. Near-Infrared Spectroscopy assessments of regional cerebral oxygen saturation for the

prediction of clinical outcomes in patients with cardiac arrest: A review of clinical impact, evolution, and future directions. Front Med

(Lausanne). 2020; 29;7:587930.

[3] Lee KS, Kim SJ, Kim DC Et al. Machine learning-based prediction of cerebral oxygen saturation based on multi-modal cerebral oximetry data. Health Informatics J. 2024;30(2):14604582241259341.

[4] Zhang CY, Yang YS, Pei MQ et al. The Association of Cerebral Oxygen Desaturation with Postoperative Cognitive Dysfunction in

Older Patients: A Review. Clin Interv Aging. 2024;19:1067-1078.

[5] Mutsuyoshi Y, Ito K, Ookawara S et al. Effects to cerebral oxygenation by arteriovenous fistula creation in patients with chronic kidney

disease. J Vasc Access. 2024;11297298241257431.

[6] Zhong W, Ji Z, Sun C. A review of monitoring methods for cerebral blood oxygen saturation. Healthcare 2021; 9: 1104.

[7] Benni PB, MacLeod D, Ikeda K et al. A validation method for near-infrared spectroscopy based tissue oximeters for cerebral and somatic tissue oxygen saturation measurements. J Clin Monit Comput. 2018;32(2):269-284.

[8] Jo YY, Shim JK, Soh S et al. Association between Cerebral Oxygen Saturation with Outcome in Cardiac Surgery: Brain as an Index Organ. J Clin Med. 2020 Mar 19;9(3):840.

[9] Bickler PE, Feiner JR, Rollins MD. Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers. Anesth Analg. 2013;117(4):813-823.

[10] MacLeod DB, Ikeda K, Vacchiano C et al. Development and validation of a cerebral oximeter capable of absolute accuracy. J Cardiothorac Vasc Anesth. 2012;26:1007–14.




DOI: http://dx.doi.org/10.70711/pmr.v3i11.10330

Refbacks

  • There are currently no refbacks.